Trends
Where:
●
Trends (
Trends
) =
Trends
area title
●
2 hours, 4 hours, 8 hours, 12 hours, 24 hours (
X hours
) = trend time interval
selections
●
SpO2% (
SpO2%
) = pulse oximetry label and unit of measure
●
PR/min (
PR/min
) = pulse rate label and unit of measure
●
RR/min (
RR/min
)= respiration rate label and unit of measure
●
Numeric values 88, 90, and 18 =
SpO2
,
PR
, and
RR
trended values at the
selected time
●
Vertical axis = trend scales
●
Horizontal axis = time (HH:MM), where the time selected with the cursor is
highlighted
About trends on the Hub
Trends show the average values of a patient’s physiological parameter data for each
minute. For this reason, trends may not always show all changes in the patient’s data,
even when that data is related to physiological alarms. The physiological parameter
trends from the past four hours may be viewed in the Hub
TRENDS
view. The trends
are updated each minute with the most recent trend value shown on the right and the
oldest on the left. The trend scale depends on the selected alarm limits; therefore, the
trends will also be updated if the alarm limits are changed on the Central Viewer.
Viewing trends on the Hub
1.
Unlock the Hub screen if it is locked.
The
LIVE
view opens.
2.
Select
TRENDS
from the toolbar.
The trend view opens. It can contain one or more parameters, depending on
the measurement setup.
5815148-01
Portrait
™
Mobile Monitoring Solution
109
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...