Alarm messages
Physiological messages
Alarm message
Location
Possible explanations
Suggested actions
Hub:
●
PR high
●
PR low
Central Viewer:
●
PR high
●
PR low
●
al. area
●
data
Measurement values are equal
to or outside the alarm limits.
●
Check the patient.
●
Adjust alarm limits on the
Central Viewer if necessary.
Hub:
●
Pulse search
●
al. area
SpO
2
and PR values are not yet
available due to ongoing signal
control and data acquisition.
●
Reposition the probe.
●
Clean the application site
(remove nail polish, for
instance).
●
If the problem persists,
contact qualified service
personnel.
Hub:
●
SpO2 artifact
●
Artifact
Central Viewer:
●
SpO2 artifact
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Artifact detected.
●
Check the patient.
●
Check the sensor and sensor
positioning.
Hub:
●
SpO2 high
●
SpO2 low
Central Viewer:
●
SpO2 high
●
SpO2 low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient.
●
Adjust alarm limits on the
Central Viewer if necessary.
Hub:
●
SpO2 critically low
Central Viewer:
●
SpO2 critically low
●
al. area
Measurement values are equal
to or outside the critical alarm
limit.
If the
SpO2 critically low
alarm
is triggered before the
SpO2
low
alarm delay has expired,
these two alarms are triggered
simultaneously.
●
Check the patient.
●
Adjust alarm limits on the
Central Viewer if necessary.
Hub:
●
SpO2 low perfusion
●
Low perfusion
Central Viewer:
●
Low perfusion
Hub:
●
al. area
●
data
Central
Viewer:
●
al. area
Low perfusion at the
measurement site.
●
Check the sensor and sensor
positioning.
●
Check the measurement site
(for instance, is the patient's
hand or finger cold).
140
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...