Impedance respiration
6.
Pair the sensor and Hub.
If the patient has a pacemaker and its status is either known
or unknown
, a notification similar to the following safety statement appears on the Hub:
WARNING
PATIENT SAFETY. Implantable Minute Ventilation Rate
Responsive Pacemakers may react with the respiration
sensor and cause the pacemaker to execute pacing
with the maximum programmed rate. Before pairing
the sensor and Hub, always check that the patient does
not have this type of pacemaker. Failing to do so may
compromise patient safety.
You can close the notification from the X icon on the upper right corner of the
display.
7.
Check the measurement.
Checking the respiration measurement
Perform the following checks to ensure that the measurement is working.
1.
On the Hub:
a. Check that the wireless network connection LED on the Hub is lit and keeps
alternating between dimming and brightening green.
b. Check that the learning animation (three dots in a horizontal line) and the
message
Learning
are displayed. When they disappear, you can see the
measurement parameter (RR) on the Hub screen.
c. Check the measurement value and the signal quality indicator to ensure
measurement quality.
2.
On the Central Viewer:
a. Check that the parameter waveform and signal quality indicator are displayed
in the Single Patient View and the parameter values in the Single and Multiple
Patient Views.
Removing respiration electrodes
1.
Support the skin at the peel line when removing the electrodes.
2.
Remove the electrodes slowly.
3.
Wipe the electrode gel from the skin with dry gauze.
Troubleshooting the measurement
Refer to the troubleshooting chapter in this manual for guidance regarding the most
common measurement-related problems. You can find a list of error messages and
alarms in the messages chapter.
98
Portrait
™
Mobile Monitoring Solution
5815148-01
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...