Managing patients and monitoring
Admitting a patient
To start monitoring, you must first admit the patient to the Central Viewer. Admitting
assigns a patient to a patient slot.
1.
Select an available patient slot in the Multiple Patient View. Available patient slots
display
Admit Patient
.
The
Admit Patient
window appears. The room and bed number of the selected
patient slot appear in the header. To change the room and bed, close the window
and select the slot assigned to the desired room and bed.
2.
Complete the
Admit Patient
information.
●
First name
: Maximum length 30 characters.
●
Last name
: Maximum length 30 characters.
●
Patient ID 1
: A site-specific patient identifier. Consult your site policy regarding
the use of this field. The name of the
Patient ID 1
field can be changed during
the system configuration.
●
Patient ID 2
: A site-specific patient identifier. Consult your site policy regarding
the use of this field. The name of the
Patient ID 2
field can be changed during
the system configuration.
●
Date of birth
: The date of birth of the patient. Enter the day and the year,
and select the month from the drop-down list. A default alarm profile is
automatically selected based on the date of birth entered.
●
Sex
: The gender of the patient.
●
Profile
: The alarm profile to use for this patient. If needed, you can change or
select a profile here. If the calculated age of the patient is less than three years
old, a warning message is displayed and a profile is not chosen automatically
because the monitoring system is not intended for patients younger than three
years old and weighing less than 10 kg. Once the
Admit Patient
information is
submitted, the alarm profile cannot be changed.
WARNING
INCORRECT ALARM PROFILE. Make sure that you enter
the correct date of birth for the patient to ensure a
correct alarm profile. A wrong alarm profile may result
in false or missing alarms.
●
Pacemaker
: The pacemaker indication for the patient. If
Yes
is selected, a
pacemaker indicator
will display with the patient data.
5815148-01
Portrait
™
Mobile Monitoring Solution
63
Содержание Portrait Hub01
Страница 12: ...12 Portrait Mobile Monitoring Solution 5815148 01...
Страница 14: ...Legal manufacturer information 14 Portrait Mobile Monitoring Solution 5815148 01...
Страница 40: ...Safety and intended use 40 Portrait Mobile Monitoring Solution 5815148 01...
Страница 72: ...Managing patients and monitoring 72 Portrait Mobile Monitoring Solution 5815148 01...
Страница 80: ...Battery management 80 Portrait Mobile Monitoring Solution 5815148 01...
Страница 82: ...Hub and modes 82 Portrait Mobile Monitoring Solution 5815148 01...
Страница 90: ...Alarms 90 Portrait Mobile Monitoring Solution 5815148 01...
Страница 112: ...Trends 112 Portrait Mobile Monitoring Solution 5815148 01...
Страница 130: ...Periodic maintenance 130 Portrait Mobile Monitoring Solution 5815148 01...
Страница 144: ...Alarm messages 144 Portrait Mobile Monitoring Solution 5815148 01...
Страница 156: ...Alarm specifications 156 Portrait Mobile Monitoring Solution 5815148 01...
Страница 164: ...Parameter specifications 164 Portrait Mobile Monitoring Solution 5815148 01...
Страница 176: ...Quick workflow overview 176 Portrait Mobile Monitoring Solution 5815148 01...
Страница 182: ...Default settings and configuration 182 Portrait Mobile Monitoring Solution 5815148 01...
Страница 186: ...Compatible devices 186 Portrait Mobile Monitoring Solution 5815148 01...
Страница 188: ...Supplies and accessories 188 Portrait Mobile Monitoring Solution 5815148 01...
Страница 197: ...content...