6.6.4.16
[SOUR]:PHAS:OUTP[:STATe](?)
<CPD>{OFF|ON|0|1}
Purpose
The 9100 AC Voltage or AC Current output phase can be used as a reference-phase output via the 'PHASE LOCK OUT' plug on
the rear panel. This command switches the reference phase on or off.
<CPD>
The 'character program data' switches the 9100: either to output the reference phase (
ON
or
1
); or not (
OFF
or
0
).
Example:
An 'AC Voltage' function output phase angle could be programmed by:
FUNC:{SIN|IMP|TRI|TRAP|SYMS};:VOLT <DNPD>;:FREQ <DNPD>;:PHAS:OUTP ON
Function Change,
∗
RST
Common Command and Power-up
The 9100 switches reference phase off (equivalent to
PHAS:OUTP {OFF
|
0}
). For details of operation and parameter limitations,
refer to Volume 1, Section 4, Sub-Sections 4.4 and 4.6 (AC Voltage and AC Current).
Response to Query Version
The instrument will return the present 'phase output state' setting:
{OFF|ON}
.