Section 8: Model 9100 — Routine Maintenance and Test
8-13
Final Width = 215mm
TODAYS DATE
TIME
TODAYS DATE
TIME
Continued Overleaf
8.3.2.2
Selftest Runs to Completion
If the selftest is not aborted, it will run to completion, and if the test is successful with
no failures, the following screen will appear for either Fast and Full test:
E X I T
Selftest.
Test completed with no failures.
P R I N T
If failures were encountered during the test, the 9100 will display the following screen:
E X I T
Selftest.
Selected test has FAILED.
Number of failures: XX
Use the softkeys to view the results.
P R I N T
VIEW
FAILS
EXIT
returns to the '
Select selftest
' menu screen.
prints out the results of the test, whether Full or Fast.
Refer to Sub-section 8.4.
If failures were encountered,
EXIT
and
remain available. Also, the number of
failures is shown on the screen, and an extra selection will be available:
VIEW FAILS
sets up a special screen for detailing the parameters of the failures
encountered (described later in Paras 8.3.2.3).