10.2-12
Section 10: Calibrating the Model 9100: Calibration Mode
Final Width = 215mm
10.2.6
Standard Calibration (STD CAL)
10.2.6.1
Function Selection
Pressing the '
STD CAL
' screen softkey displays the DC Voltage CAL mode function
screen shown below, which always appears when you first enter Standard Calibration.
If you wish to calibrate other functions which are available for user-calibration, they can
be selected in the same way as for normal operation — i.e. Voltage, Current and
Resistance are selected simply by pressing their corresponding front-panel push-button,
AC Voltage is selected from the DC Voltage function screen, and Capacitance is selected
via the Aux push-button and Auxiliary Function menu screen (see Section 4 of the Model
9100 Universal Calibration System User's Handbook - Volume 1 - Operation.)
Note that the Model 9100's Frequency, Conductance, Temperature and Logic functions
cannot be externally calibrated (they are either derived directly from functions that can
be calibrated or are calibrated 'for life' during manufacture). Attempting to select these
functions while the Model 9100 is in CAL mode will result in an error message being
displayed:-
No calibration for this function
V
t i m e a n d d a t e
+
.
Z E R O
±
V
x
÷
V
V
TARGET
C A L
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......