Section 10: Calibrating the Model 9100: Calibration Mode
10.2-11
Final Width = 215mm
sHi
sLo
Hi
Lo
I
+
I-
!
1500Vpk
max
15Vpk
max
15Vpk
max
HIGH VOLTAGE
DANGER
!
Resistance Temperature Detector,
or
DMM on 100
Ω
/1k
Ω
Ohms Range.
Use low exitation current (1mA ±20%)
CJC Pod Module, shown plugged into the D-type socket on the
9100 front panel.
(Note that if the 9105 lead set is in use, the calibration can be
performed with the module plugged into the D-type socket on the
end of the connection box underneath the work mat.)
PRT fully inserted
into the Hole in the
end of the Pod
I+
I-
Hi
Lo
4-wire connection.
Do not use 2-wire
connection as errors
will be excessive.
Fig. 10.2.5.1 Setup for Cold Junction Calibration