7-12
Section 7: Model 9100: Specifications
Final Width = 215mm
7.6.7
AC Current Accuracy (Triangular-Wave)
Frequency
Output Current Span
Accuracy †
Voltage
Band
[2]
RMS
Peak
±
(% Floor)
Compliance
(Hz)
1 Year — Tcal
[1]
±
5
°
C
(VRMS)
10 - 1k
0
µ
A - 26.12
µ
A
0
µ
A - 45.25
µ
A
0.21 + 1.8
µ
A
3.2
10 - 1k
26.12
µ
A - 2.612mA
45.25
µ
A - 4.525mA
0.21 + 0.6
µ
A
3.2
10 - 1k
2.612mA - 26.12mA 4.525mA - 45.25mA
0.21 + 6.4
µ
A
3.2
10 - 1k
26.12mA - 261.2mA 45.25mA - 452.5mA
0.24 + 64
µ
A
3.2
10 - 100
0.2612A - 2.612A
0.4525A - 4.525A
0.30 + 960
µ
A
2.0
10 - 100
2.612A - 16.30A*
4.525A - 28.23A
0.40 + 13.8mA
2.0
♦
10 - 65
2.612A - 26.12A
♥
4.525A - 45.25A
0.40 + 16.8mA
2.0
10 - 65
26.12A - 163.0A*
♥
45.25A - 282.3A
0.60 + 162mA
2.0
♦
10 - 65
13.06A - 130.6A
♣
22.62A - 226.2A
0.40 + 84mA
2.0
10 - 65
130.6A - 815.0A*
♣
226.2A - 1411A
0.60 + 0.82A
2.0
♦
Ratios and Factors based on
Triangular-Wave Unit-Value Peak
Peak:
1.0000
Peak to Peak:
2.0000
RMS:
0..5774
Mean:
0.5000
Crest Factor:
1.7321
Form Factor:
1.1547
∗
=
With output 'ON', maximum duty cycle of (>0.644FS :
≤
0.644FS) is (1 : 4).
Continuous output >0.644FS will automatically reduce to <0.644FS after 2 Minutes.
† =
Total uncertainty includes compliance errors for Voltage
≤
0.5VRMS.
Above 0.5V, add appropriate compliance error, except for Outputs marked
♥
and
♣
.
♥
=
Refers to accuracy at 9100 output terminals. With Option 200 10 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♣
=
Refers to accuracy at 9100 output terminals. With Option 200 50 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♦
=
For frequencies <40Hz, compliance voltage is reduced by 0.5V RMS.
7.6
AC Current Specifications
(Contd.)
7.6.8
AC Current Accuracy (Trapezoidal-Wave)
Frequency
Output Current Span
Accuracy †
Voltage
Band
[2]
RMS
Peak
±
(% Floor)
Compliance
(Hz)
1 Year — Tcal
[1]
±
5
°
C
(VRMS)
10 - 1k
0
µ
A - 39.91
µ
A
0V - 45.25
µ
A
0.21 + 1.8
µ
A
4.0
10 - 1k
39.91
µ
A - 3.991mA
45.25
µ
A - 4.525mA
0.21 + 0.6
µ
A
4.0
10 - 1k
3.991mA - 39.91mA 4.525mA - 45.25mA
0.21 + 6.4
µ
A
4.0
10 - 1k
39.91mA - 399.1mA 45.25mA - 452.5mA
0.24 + 64
µ
A
4.0
10 - 100
0.3991A - 3.200A
0.4525A - 3.628A
0.30 + 960
µ
A
2.3
10 - 100
3.200A - 19.20A*
3.628A - 21.77A
0.40 + 13.8mA
2.3
♦
10 - 65
3.991A - 32.00A
♥
4.525A - 36.28A
0.40 + 16.8mA
2.3
10 - 65
32.00A - 192.0A*
♥
36.28A - 217.7A
0.60 + 162mA
2.3
♦
10 - 65
19.95A - 160.0A
♣
22.62A - 181.4A
0.40 + 84mA
2.3
10 - 65
160.0A - 960.0A*
♣
181.4A - 1088A
0.60 + 0.82A
2.3
♦
Ratios and Factors based on
Trapezoidal-Wave Unit-Value Peak
Peak:
1.0000
Peak to Peak:
2.0000
RMS:
0.8819
Mean:
0.8333
Crest Factor:
1.3389
Form Factor:
1.0583
∗
=
With output 'ON', maximum duty cycle of (>0.547FS :
≤
0.547FS) is (1 : 4).
Continuous output >0.547FS will automatically reduce to <0.547FS after 2 Minutes.
† =
Total uncertainty includes compliance errors for Voltage
≤
0.5VRMS.
Above 0.5V, add appropriate compliance error, except for Outputs marked
♥
and
♣
.
♥
=
Refers to accuracy at 9100 output terminals. With Option 200 10 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♣
=
Refers to accuracy at 9100 output terminals. With Option 200 50 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♦
=
For frequencies <40Hz, compliance voltage is reduced by 0.5V RMS.
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......