Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
10.3-5
Final Width = 215mm
TODAYS DATE
TIME
10.3.3.1
The Target Selection Screen
Once the correct hardware configuration has been established by setting the 9100 output
control display to a suitable output value, pressing the TARGET softkey will present a
'target selection' screen similar to that shown below. This screen shows those target
values at which the selected hardware configuration was last calibrated, indicated by the
"SAVED CALIBRATION TARGETS" title positioned above these target values.
10.3.3
Selecting Target Calibration Values
The currently-selected output value (and, in the case of AC Voltage or AC Current the
frequency value) is compressed to smaller characters, and the digit cursor is deactivated
so that these values cannot be changed while this screen is presented.
Depending on the selected function and hardware configuration, the screen can display
up to four target values, numbered sequentially from top to bottom. Also, the screen will
show the same number of 'Factor' keys, labelled 'Factor 1', 'Factor 2', 'Factor 3' etc.,
plus an additional screen key labelled 'DEFLT'
V
V
C A L
+
1 .
V
TARGET
F a c t o r
1
F a c t o r
2
DEFLT
S A V E D C A L I B R A T I O N T A R G E T S
1
→
— 3 . 5 4 2 6
l o w
2
→
+ 3 . 5 4 2 6
h i g h
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......