Section 6: 9100 System Operation — SCPI Language
6-57
Final Width = 215mm
6.6.6.5
SYST:SVOL (?)
<DNPD>
Purpose
This command sets the voltage value of the threshold of operation for the High Voltage Warning as employed in DC Voltage and
AC Voltage functions. The 9100 need not be currently set in either of these functions to program the voltage.
<DNPD>
The 'decimal numeric program data' is a number which sets the required voltage safety warning threshold, expressed in units of DC
or RMS AC Volts. It should be unsigned. The parameter
<DNPD>
must have a value in the range 10.000V to 110.000V inclusive.
Response to Query Version:
SYST:SVOL?
The instrument will return the present DC or AC voltage safety warning threshold value. The returned number will be in standard
scientific unsigned format (for example: 90V would be returned as
9.0E1
).
6.6.6.6
SYST:VERS?
Purpose
This query returns a numeric value corresponding to the SCPI version number for which the 9100 complies.
Response
SYST:VERS?
returns an <Nr2> formatted numeric value corresponding to the SCPI version number for which the 9100 complies.
At the time of writing, this will be 1994.0.
6.6.6.7
SYST:FOR?
Purpose
This query returns the present date format, as programmed locally.
Response
SYST:FOR?
returns one of three sets of three characters:
DMY
,
MDY
or
YMD (Day/Month/Year
,
Month/Day/Year
or
Year/Month/
Day).
Formatting is carried out from a screen in 'Configuration' mode (Refer to Volume 1 of this User's Handbook; Section 3,
Subsection 3.3.2, paras 3.3.2.10).
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......