Section 6: 9100 System Operation — SCPI Language
6-29
Final Width = 215mm
6.6.3.4
OUTP:ISEL(?)
<CPD>{HIGHi|HI50|HI10|LOWi}
Purpose
The I+ Output in Current function can be passed out via the unguarded front panel I+ terminal, or on a guarded line (pin 8) through
the D-type socket ('Signal Output' J109) beneath the terminals. Pin 7 of J109 is connected to internal Guard shields for this purpose,
and is available for external guard connection. The return is always via the front panel main I- terminal.
The front panel I+ terminal can carry all DC currents and AC currents up to the maximum -20A to +20A output (refer to
Volume 1, Section 4, paras 4.5.5.3 for time limitations). Pin 8 of J109 is limited to -1A to +1A output.
OUTP:ISEL HIGH
switches current outputs via the front panel I+ terminal,
OUTP:ISEL HI50
(Option 200) switches current outputs via the front panel I+ terminal, and provides equivalent
currents of
±
(16.0A to 1000A) into the 50-turn current coil.
OUTP:ISEL HI10
(Option 200) switches current outputs via the front panel I+ terminal, and provides equivalent
currents of
±
(3.2A to 200A) into the 10-turn current coil.
OUTP:ISEL LOW
switches current outputs via J109 pin 8.
Response to Query Version
The 9100 will return
HIGH/HI50/HI10/LOW
to match the active programming parameter.
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......