10.4-26
Section 10: Calibrating the Model 9100: DC Current Function
Final Width = 215mm
Table 10.4.5.3: Auxiliary DC Current Calibration Points
Hardware Spans
Calibration Targets
Default
Minimum
Maximum
0.000mA to
±
320.0mA
-300.000mA
-320.000mA
-180.00mA
(one range)
+300.000mA
+180.000mA
+320.000mA
±
0.321A to
±
3.200A
-3.00000A
-3.20000A
-1.80000A
+3.00000A
+1.80000A
+3.20000A
±
3.21 to
±
32.00A
-10.0000A
-10.5000A
-7.00000A
+10.0000A
+7.00000A
+10.5000A