6-28
Section 6: 9100 System Operation — SCPI Language
Final Width = 215mm
6.6.3
OUTPut Subsystem
This subsystem is used to select the output connections of the 9100, switch the output on and off, and switch the lead compensation
on and off.
6.6.3.1
OUTPut Subsystem Table
Keyword
Parameter Form
Notes
OUTPut
[:STATe](?)
<CPD>{ON|OFF|0|1}
:COMPensation(?)
<CPD>{ON|OFF|0|1}
:ISELection(?)
<CPD>{HIGHi|HI50turn|HI10turn|LOWi}
6.6.3.2
OUTP[:STAT](?) <CPD>{ON|OFF|0|1}
Purpose
This command turns the 9100 output on and off.
•
ON
or
1
will set the output on
•
OFF
or
0
will set the output off
Response to Query Version
The 9100 will return
ON
if output is on, or
OFF
if output is off.
6.6.3.3
OUTP:COMP(?)
<CPD>{ON|OFF|0|1}
Purpose
This command switches the output connections for 4-wire/2-wire in the impedance functions.
Impedance Functions:
Resistance, Conductance, Capacitance and PRT Temperature.
Response to Query Version
The 9100 will return
ON
if compensation is on (4-wire), or
OFF
if compensation is off (2-wire).