10.4-20
Section 10: Calibrating the Model 9100: DC Current Function
Final Width = 215mm
10.4.5
DC Current Calibration
10.4.5.1
Introduction
This section is a guide to calibrating the Model 9100's DC Current Function using its front
panel controls. The following topics are covered:
10.4.5.2
Calibration Equipment Requirements
10.4.5.3
Interconnections
10.4.5.4
Calibration Setup
10.4.5.5
Calibration Procedure
10.4.5.2
Calibration Equipment Requirements
A traceably characterized, long scale-length, Standards DMM connected to the 9100
output terminals by short,
high-quality leads. For example,
a
Fluke
Model 1281
Digital Multimeter. Note that some DMM's may require a precision current shunt to
measure up to 10A.