10.4-10
Section 10: Calibrating the Model 9100: DC Voltage Function
Final Width = 215mm
10.4.3.6
Calibration of Auxiliary DC Voltage (for Power Option)
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.3.2, and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select Power and DC Volts by pressing the front-
panel AUX key followed by the softkey sequence
POWER
*
,
W
, ,
VOLTAGE
.
Refer also to Fig. 10.2.2.
*
This keystroke is only required if one of the oscilloscope calibration options (Option 250 or Option 600) is
fitted.
AUX
CHANNEL
MODE