Section 6: 9100 System Operation — SCPI Language
6-49
Final Width = 215mm
6.6.4.26
[SOUR]:CONT[:LEVE][:IMM][:AMPL](?)
<DNPD>
Purpose
This command activates the Continuity function. The value
<DNPD>
is the desired resistance value in ohms demanded by the user.
6.6.4.27
[SOUR]:CONT:TCUR?
Purpose
This query only command requests the test current provided by the UUT for confirmation that the UUT is providing sufficient test
current to comply with its calibration specification (also see Note 1 below).
6.6.4.28
[SOUR]:INS[:LEVE][:IMM][:AMPL](?)
<DNPD>
Purpose
This command activates the Insulation Resistance function. The value
<DNPD>
is the resistance value in ohms demanded by the
user.
6.6.4.29
[SOUR]:INS:UUT_I(?)
<CPD>
Purpose
This command operates in the same manner as that for Active Resistance, with the exception that the only two permissible values
of current setting
<CPD>
are
HIGH
and
SUPer
.
6.6.4.30
[SOUR]:INS:TVOL?
Purpose
This query only command returns the value of UUT test voltage (also see Note 1 below).
6.6.4.31
[SOUR]:INS:TCUR?
Purpose
This query only command returns the value of UUT test current (also see Note 1 below).
Note 1 : If the commands
:CONT:TCUR?
,
:INS:TVOL?
or
:INS:TCUR?
are executed when the output is OFF, or the Model
9100 is not in the relevant function, the value returned will be
2.0E+35
.
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......