10.4-12
Section 10: Calibrating the Model 9100: DC Voltage Function
Final Width = 215mm
Hardware Spans
Calibration Targets
Default
Minimum
Maximum
0.000 to 32.00mV
-30.0000mV
-32.0000mV
-18.0000mV
+30.0000mV
+18.0000mV
+32.0000mV
32.1mV to 320.0mV
-300.000mV
-320.000mV
-180.000mV
+300.000mV
+180.000mV
+320.000mV
0.321V to 3.200V
-3.00000V
-3.20000V
-1.80000V
+3.00000V
+1.80000V
+3.20000V
3.201V to 7.500V
-6.00000V
-6.50000V
-4.00000V
+6.00000V
+4.00000V
+6.50000V
Table 10.4.3.3: Auxiliary DC Voltage Calibration Points