10.4-14
Section 10: Calibrating the Model 9100: AC Voltage Function
10.4.4.1
Introduction
This section is a guide to calibrating the Model 9100 AC Voltage Function using its front
panel controls. The following topics are covered:
10.4.4.2
Calibration Equipment Requirements
10.4.4.3
Interconnections
10.4.4.4
Calibration Setup
10.4.4.5
Calibration Procedure
10.4.4.2
Calibration Equipment Requirements
A traceably characterized, long scale-length, Standards DMM connected to the Model
9100
's
output terminals by a short,
high-quality coaxial cable.
For example,
a
Fluke
Model 1281 Digital Multimeter.