7-30
Section 7: Model 9100: Specifications
Final Width = 215mm
7.19.9 Phase Accuracy (Sinusoidal Waveshape)
Voltage Output
Frequency
Output Phase
Band (Hz)
Uncertainty (degrees)
0.30000V to 105.000V
10 - 65
0.07
°
65 - 1k
0.07 + 0.001 x (f - 65)
°
105.001V to 750.000V
[1]
45 - 65
0.16
°
65 - 1k
0.16 + 0.0037 x (f - 65)
°
Current Output
Frequency
Output Phase
Band (Hz)
Uncertainty (degrees)
[2]
0.00000A to 20.0000A
10 - 65
0.08
°
65 - 1k
0.08 + 0.0008 x (f - 65)
°
>1k
0.08 + 0.0012 x (f - 65)
°
03.2000A to 1000.00A
[3]
10 - 65
0.23
°
65 - 440
0.23 + 0.003 x (f - 65)
°
Auxiliary Channel Voltage Output
Frequency
Output Phase
Band (Hz)
Uncertainty (degrees)
[2]
0.32mV to 7.500A
10 - 65
0.07
°
65 - 1k
0.07 + 0.001 x (f - 65)
°
>1k
0.07 + 0.0015 x (f - 65)
°
NOTES: [1]
Maximum load current: 2mA. Maximum load capacitance: 200pF
[2]
To obtain the Output Phase Uncertainty with respect to PHASE LOCK OUT or PHASE LOCK IN when operating in POWER
mode, add 0.07
°
.
[3]
With the 10-turn or 50-turn coil output selected (Option 200).