Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
10.3-7
Final Width = 215mm
10.3.3.3
Using the Default Calibration Targets
As mentioned earlier, the Model 9100's firmware contains a complete set of recommended
target calibration values, for every hardware configuration of every function that can be
directly calibrated. If you wish to use these 'default' values, rather than the values at which
your 9100 was previously calibrated (assuming that the 'saved' and 'default' values are
different), simply press the DEFLT (Default) softkey which appears in the target
selection screen. The Default Calibration Targets for the selected hardware configuration
will be displayed as shown below. Re-pressing the DEFLT will take you back to the
display of Saved Calibration Targets — i.e. the DEFLT key toggles between the
'Saved' and 'Default' sets of target values.
To select one of the displayed 'default' target values for calibration, press the corresponding
Factor softkey (i.e. the 'Factor 1' key to calibrate at target value one, the 'Factor 2' key
to calibrate at target value two. etc..)
This will transfer you to a Calibrate screen, in which you can determine the output error
and correct it — see Section 10.3.4.
TODAYS DATE
TIME
V
V
C A L
+
1 .
V
TARGET
F a c t o r
1
F a c t o r
2
D E F A U L T C A L I B R A T I O N F A C T O R S
1
→
— 3 .
l o w
2
→
+ 3 .
h i g h
DEFLT
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......