10.4-6
Section 10: Calibrating the Model 9100: DC Voltage Function
Final Width = 215mm
10.4.3.1
Introduction
This section is a guide to calibrating the Model 9100's DC Voltage Function using its front
panel controls. The following topics are covered:
10.4.3.2
Calibration Equipment Requirements
10.4.3.3
Interconnections
10.4.3.4
Calibration Setup
10.4.3.5
Calibration Procedure
10.4.3.2
Calibration Equipment Requirements
A traceably characterized, long scale-length, Standards DMM connected to the Model
9100
's
output terminals by short,
high-quality leads. For
example,
a
Fluke
Model
1281
Digital Multimeter.
10.4.3
DC Voltage Calibration