10.3-16
Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
Final Width = 215mm
TODAYS DATE
TIME
10.3.6
Exit from Calibration - Cal Date and Cal Due Date
Once all calibration has been completed, you will wish to return to normal operation,
which requires a short 'exit' process. The normal means of exit from Calibration mode
is to press the
Mode
key on the right of the front panel. When you do this, the 9100 will
present a Warning screen to indicate that the 9100 calibration may have changed, and
to offer you the following options:
•
altering the date-stamp on any directly-printed results certificate (or not),
•
entering or altering the calibration due date,
•
indicate the advance warning period required before the cal due date.
The following paragraphs show the screens, required actions and consequences.
10.3.6.1
Exit: Mode Key — Warning Screen
To exit from Calibration, press the
Mode
key on the right of the front panel. The
9100 will present a 'Warning' display on
the screen:
TODAYS DATE
TIME
C A L D U E
D A T E
WARNING
The calibration of the 9100
may now have changed, if you
wish to date stamp the cal or
alter the cal due date, select
one of the softkeys below
The stored cal date is:
25/01/1995
The stored due date is:
25/04/1995
E X I T
C A L
D A T E
10.3.6.2
Exit Only
If you do not wish to alter the date-stamp, and do not wish to alter the cal due date or its
advance warning period, press the
EXIT
key. This terminates the calibration session, and
you will be returned to the Mode Selection screen to select another mode.
10.3.6.3
Update the Date Stamp on a Certificate
If you wish to update the date stamp to
today's date, press the
CAL DATE
key.
The 9100 presents the Cal Date screen
→
The alpha-numeric keypad is locked out.
If updating the date-stamp is all that is
required, without altering the due date,
press the
EXIT
screen key to return to the
Mode Selection screen.
CAL DATE
The calibration date of
the 9100 has now been
set to
25/01/1995
E X I T
C A L D U E
D A T E
Содержание 9100 Series
Страница 1: ...Final Width 215mm USER S HANDBOOK Model 9100 Universal Calibration System Volume 2 Performance ...
Страница 70: ......
Страница 76: ......
Страница 94: ......
Страница 131: ...Section 7 Model 9100 Specifications 7 35 Final Width 215mm ...
Страница 132: ......
Страница 160: ......
Страница 172: ......
Страница 174: ......
Страница 190: ......
Страница 208: ......
Страница 272: ......
Страница 273: ......
Страница 274: ......