6-30
Section 6: 9100 System Operation — SCPI Language
Final Width = 215mm
6.6.4
SOURce Subsystem
This subsystem is used to select the sources of 9100 output .
6.6.4.1
SOURce Subsystem Table
Keyword
Parameter Form
Notes
[SOURce]
:FUNCtion
[:SHAPe](?)
<CPD>{DC|SINusoid|PULSe|SQUare|IMPulse|TRIangle|TRAPezoid|SYMSquare}
:VOLTage
[:LEVEl]
[:IMMediate]
[:AMPLitude](?)
<DNPD>
:HIGH(?)
<DNPD>
:LOW(?)
<DNPD>
:CURRent
[:LEVEl]
[:IMMediate]
[:AMPLitude](?)
<DNPD>
:RESistance
[:LEVEl]
[:IMMediate]
[:AMPLitude](?)
<DNPD>
:UUT_I(?)
<CPD>{LOW|HIGH|SUPer}
:CONDuctance
[:LEVEl]
[:IMMediate]
[:AMPLitude](?)
<DNPD>
:UUT_I(?)
<CPD>{LOW|HIGH|SUPer}
:CAPacitance
[:LEVEl]
[:IMMediate]
[:AMPLitude](?)
<DNPD>
:UUT_I(?)
<CPD>{LOW|SUPer}
:FREQuency
[:CW|FIXed](?)
<DNPD>
:PHASe
[:ADJust](?)
<DNPD>
:INPut
[:STATe](?)
<CPD>{ON|OFF|0|1}
:OUTPut
[:STATe](?)
<CPD>{ON|OFF|0|1}
Содержание 9100 Series