7-26
Section 7: Model 9100: Specifications
Final Width = 215mm
7.18.3 Continuity Resistance Accuracy and Resolution (4-wire connection)
Resistance Output
Accuracy
Absolute
±
(% of Floor)
Resolution
1Year — Tcal
±
5
°
C
[1]
00.0000
Ω
to 40.0000
Ω
0.1 + 50m
Ω
0.1m
Ω
040.001
Ω
to 400.000
Ω
0.035 + 100m
Ω
1m
Ω
0.40001 k
Ω
to 4.00000 k
Ω
0.035 + 200m
Ω
10m
Ω
Hardware Configuration Limits
Source Current Limits
on Span of Output Resistance
00.0000
Ω
to 40.0000
Ω
25mA to 350mA
040.001
Ω
to 400.000
Ω
5mA to 70mA
0.40001 k
Ω
to 4.00000 k
Ω
500
µ
A to 7mA
7.18.4 Continuity Resistance Source Current — Limits
7.18.5 Other Continuity Resistance Specifications
Maximum Measurement Voltage:
10V
Current Measurement:
Range:
100
µ
A to 350mA
Accuracy:
±
1%
7.18 Insulation/Continuity Specifications