Micropilot FMR56, FMR57 FOUNDATION Fieldbus
Operating menu
Hauser
157
• Digital output 2
• Digital output 3
• Digital output 4
• Digital output 5
• Digital output 6
• Digital output 7
• Digital output 8
Factory setting
Off
Additional information
The
Digital output AD 1
and
Digital output AD 2
options refer to the Advanced
Diagnostic Blocks. A switch signal generated in these blocks can be transmitted via the
switch output.
Assign limit
Navigation
Setup → Advanced setup → Switch output → Assign limit
Prerequisite
Switch output function (→ 156)
=
Limit
Description
Select process variable for limit monitoring.
Selection
• Off
• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Relative echo amplitude
• Area of incoupling
Factory setting
Off
Assign diagnostic behavior
Navigation
Setup → Advanced setup → Switch output → Assign diag. beh
Prerequisite
Switch output function (→ 156)
=
Diagnostic behavior
Description
Select diagnostic behavior for switch output.
Selection
• Alarm
• Alarm or warning
• Warning
Factory setting
Alarm
Содержание Micropilot FMR56