Mass
Spectrometer
PlasmaQuant MS Series
150
8.7.3
Mass analyzer Assembly
Figure 8-39 - Mass analyzer assembly exploded diagram
1.
Vent the vacuum.
2.
Turn off the instrument power.
3.
Remove the front and top covers as well as the top plate covering the analyzer
assembly.
4.
Using steps from section 8.7.2, page 147 remove the Detector Assembly from
the instrument.
5.
With the detector assembly removed, loosen the two screws holding the
quadrupole supply tabs onto the quadrupole assembly using a clean flat blade
screwdriver.
6.
Remove the 4 bolts located on top of the mass analyzer assembly holding it to
VC3.
7.
Carefully extract the mass analyzer assembly from the top of the instrument.
8.
To replace assembly, follow steps in reverse.
Mass Analyzer Assembly
Removal and
Replacement