PlasmaQuant MS Series
Power
Supply
189
11
Power Supply
11.1
Introduction
This chapter contains circuit descriptions relating to the power supply PCB’s used in the
ICPMS. For circuit diagrams relating to the Analytik Jena designed circuitry refer to the
relevant section of the Schematics manual.
WARNING
Never override or disable the instrument interlocks. Failure to observe this warning
could expose the operator to high voltage resulting in death or serious injury.
11.2
System Description
The ICP-MS power supply assemblies are listed below.
Ion Optics & Detector supply PCB
Quadrupole controller
Electrical box module
RF generator
DC power supply
Utility box
11.2.1
Ion Optics High Voltage Power Supply
The 11 ion optics supplies are generated on the Ion Optics High Voltage PCB board as
well as the detector high voltage power supply signals. It is located on back of the
instrument. Test points for the lens supplies are located from the through the cover.