PlasmaQuant MS Series
Sample
Introduction
23
4.2
Theory of Operation
The sample introduction system generates an aerosol, in a carrier gas, from a liquid
sample and directs it into the center of the toroidal plasma. As the sample passes
through the plasma, it rapidly undergoes the following transformation:
1.
Desolvation stage (drying of the sample)
2.
Molecular decomposition
3.
Formation of ions
4.2.1
Nebulizer
A pneumatic nebulizer and a stream of argon gas generate the sample aerosol. The
micro-mist glass concentric is the standard type of pneumatic nebulizer used with the
instrument.
Figure 4-2 - Standard Micro-mist Nebulizer (Part #: 418-88070-0)
Liquid sample input
Argon gas input, typical
flow of 1 L/m.