Gas
Control System
PlasmaQuant MS Series
50
5.5.4
Removing the Argon Inlet Manifold with RG1 and RG2 Attached
Figure 5-16 - Positioning of RG1 and RG2 within gas control box
1.
Remove covers as outlined in section 5.5.1, page 47 to reach gas box assembly.
2.
Loosen the two screws securing the inlet manifold on the frame (showing in the
picture above).
3.
Disconnect the argon inlet solenoid and argon pressure switch cables from the
sample introduction PWB.
4.
Remove the Swagelok fittings at the outlets (left) of the two regulators (RG1 and
RG2).
5.
Carefully remove the two regulators and inlet manifold out of the gas box
assembly.
NOTE
Clearly mark each cable/tube before removing for easy identification. To replace the
assembly, use the same procedures in reverse.