Configure process measurement
t
Example: Cutoff interaction
Configuration:
•
mA Output Process Variable
for the primary mA output:
Mass Flow Rate
•
Frequency Output Process Variable
:
Mass Flow Rate
•
AO Cutoff
for the primary mA output:
15
grams/second
•
Mass Flow Cutoff
:
10
grams/second
Result:
• If the mass flow rate drops below 15 grams/second but not below 10 grams/second:
− The primary mA output will report zero flow.
− The frequency output will report the actual flow rate, and the actual flow rate will be used
in all internal processing.
• If the mass flow rate drops below 10 grams/second, both outputs will report zero flow, and 0 will
be used in all internal processing.
5.3
Configure volume flow measurement for liquid
applications
Display
OFF-LINE MAINT→OFF-LINE CONFG→UNITS→VOL
ProLink II
ProLink→Configuration→Flow
Field Communicator
Configure→Manual Setup→Measurements→Flow
The volume flow measurement parameters control how liquid volume flow is measured and reported.
The volume flow measurement parameters include:
•
Volume Flow Type
•
Volume Flow Measurement Unit
•
Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow. You must choose one or
the other.
5.3.1
Configure
Volume Flow Type
for liquid applications
Display
OFF-LINE MAINT→OFF-LINE CONFG→VOL→VOL TYPE LIQUID
ProLink II
ProLink→Configuration→Flow→Vol Flow Type→Liquid Volume
Field Communicator
Configure→Manual Setup→Measurements→Gas Standard Volume→Volume Flow Type→Liquid
Volume Flow Type
controls whether liquid or gas standard volume flow measurement will be implemented.
Configuration and Use Manual
69
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...