Troubleshooting
Table 10-2
Status alarms and recommended actions
continued
Alarm code
Description
Cause
Recommended actions
A012
Calibration too high
Mechanical zero:
corresponds with high
forward flow — the
resulting zero was
greater than 3 μs
Temperature/Density
calibration: many
possible causes
•
Verify that there is no flow through the
sensor, then retry the procedure.
•
Cycle power to the meter, then retry the
procedure.
A013
Zero too noisy
Mechanical zero:
Unstable value exists
•
Remove or reduce sources of
electromechanical noise (e.g., pumps,
vibration, pipe stress), then retry the
procedure.
•
Cycle power to the meter, then retry the
procedure.
A014
Transmitter failed
Many possible causes
•
Cycle power to the meter.
•
Contact Micro Motion.
A016
Line temperature
out-of-range
The value computed
for the resistance
of the Line RTD is
outside limits
•
Check the sensor wiring. See Section
10.27.1 .
•
Verify process conditions.
•
Contact Micro Motion.
A017
Meter RTD
temperature
out-of-range
The value computed
for the resistance of
the Meter/Case RTD
is outside limits
•
Check the sensor wiring. See Section
10.27.1 .
•
Verify process conditions. Temperature
should be between –200
°
F and +400
°
F.
•
Verify characterization parameters.
•
Contact Micro Motion.
A018
EEPROM checksum
error
•
Cycle power to the meter.
•
Contact Micro Motion.
A019
RAM or ROM test error
•
Cycle power to the meter.
•
Contact Micro Motion.
A020
Calibration factors
unentered
The flow calibration
factor and/or K1 has
not been entered since
the last master reset.
•
Verify characterization parameters.
A021
Incorrect sensor type
The sensor is
recognized as a
straight tube but the
K1 value indicates a
curved tube, or vice
versa.
•
Verify characterization parameters.
180
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...