Chapter 4
Quick start with the Field Communicator
Topics covered in this chapter:
♦ Apply power
♦ Connect with the Field Communicator
♦ Configuration tips and tricks
♦ Configure the primary mA output to report mass flow rate in a user-selected
♦ Perform a loop test
♦ Trim mA outputs
♦ Zero the flowmeter
♦ Test or tune the system using sensor simulation
♦ Enable/disable HART security
4.1
Apply power
Prerequisites
Before you apply power to the flowmeter, close and tighten all housing covers.
To prevent ignition of flammable or combustible atmospheres, make sure all covers are
tightly closed. For hazardous area installations, applying power to the unit while housing
covers are removed or loose can cause an explosion.
Procedure
Turn on the electrical power at the power supply.
The flowmeter will automatically perform diagnostic routines. For transmitters with a display, the status
LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED
exhibits different behavior, an alarm condition is present.
Postrequisites
Allow the electronics to warm up for approximately 10 minutes before relying on process measurements.
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up
to 10 minutes to warm up completely.
4.2
Connect with the Field Communicator
Prerequisites
The following HART device description (DD) must be installed on the Field Communicator: DD v2.
A connection from the Field Communicator to your transmitter allows you to read process data, configure
the transmitter, and perform maintenance and troubleshooting tasks.
Configuration and Use Manual
49
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...