Integrate the meter with the control system
Postrequisites
Important
Whenever you change an mA output parameter, verify all other mA output parameters before returning
the flowmeter to service. In some situations, the transmitter automatically loads a set of stored values,
and these values may not be appropriate for your application.
7.1.1
Configure
mA Output Process Variable
Display
OFF-LINE MAINT→OFF-LINE CONFG→IO→AO 1→SRC
OFF-LINE MAINT→OFF-LINE CONFG→IO→AO 2→SRC
ProLink II
ProLink→Configuration→Analog Output→Primary/Secondary Output→PV/SV Is
Field Communicator
Configure→Manual Setup→Inputs/Outputs→mA Output 1→Primary Variable
Configure→Manual Setup→Inputs/Outputs→mA Output 2→Secondary Variable
mA Output Process Variable
controls the variable that is reported over the mA output.
Prerequisites
If you plan to configure an output to report volume flow, ensure that you have set
Volume Flow Type
as
desired:
Liquid
or
Gas Standard Volume
.
If you plan to configure an output to report a concentration measurement process variable, ensure that
the concentration measurement application is configured so that the desired variable is available.
If you are using the HART variables, be aware that changing the configuration of
mA Output Process
Variable
will change the configuration of the HART Primary Variable (PV) and/or the HART Secondary
Variable (SV).
Procedure
Set
mA Output Process Variable
as desired.
Default settings are as follows:
• Primary mA output:
Mass Flow Rate
• Secondary mA output:
Density
Options for
mA Output Process Variable
The transmitter provides a basic set of options for
mA Output Process Variable
, plus several
application-specific options. Different communications tools use different labels for the options.
Options for
mA Output Process Variable
are listed in Table 7-1 .
Table 7-1
Options for
mA Output Process Variable
Label
Process variable
Display
ProLink II
Field Communicator
Mass flow rate
MFLOW
Mass Flow Rate
Mass flo
Volume flow rate
VFLOW
Volume Flow Rate
Vol flo
118
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...