Chapter 2
Quick start with the display
Topics covered in this chapter:
♦ Apply power
♦ Configuration tips and tricks
♦ Configure the primary mA output to report mass flow rate in a user-selected
♦ Perform a loop test
♦ Zero the flowmeter
2.1
Apply power
Prerequisites
Before you apply power to the flowmeter, close and tighten all housing covers.
To prevent ignition of flammable or combustible atmospheres, make sure all covers are
tightly closed. For hazardous area installations, applying power to the unit while housing
covers are removed or loose can cause an explosion.
Procedure
Turn on the electrical power at the power supply.
The flowmeter will automatically perform diagnostic routines. For transmitters with a display, the status
LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED
exhibits different behavior, an alarm condition is present.
Postrequisites
Allow the electronics to warm up for approximately 10 minutes before relying on process measurements.
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up
to 10 minutes to warm up completely.
2.2
Configuration tips and tricks
Review these tips before beginning configuration.
2.2.1
Access to
OFFLINE
menu
Access to the
OFFLINE
menu may be disabled. To configure the transmitter using the display, you must
enable access to the
OFFLINE
menu.
Configuration and Use Manual
19
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...