Default values and ranges
Table A-1
Transmitter default values and ranges
continued
Type
Setting
Default
Range
Comments
Base mass unit
g
Base mass time
sec
Mass flow conversion factor
1
Base volume unit
L
Base volume time
sec
Special units
Volume flow conversion factor
1
Primary variable
Mass flow
Secondary variable
Density
Tertiary variable
Mass flow
Variable
mapping
Quaternary variable
Volume flow
Primary variable
Mass flow
LRV
–200.00000
g/s
URV
200.00000 g/s
AO cutoff
0.00000 g/s
AO added damping
0.00000 sec
LSL
–200 g/s
USL
200 g/s
Read-only.
LSL and USL are calculated
based on the sensor size and
characterization parameters
MinSpan
0.3 g/s
Read-only
Fault action
Downscale
AO fault level – downscale
2.0 mA
0.0 – 3.6 mA
AO fault level – upscale
22 mA
21.0 – 24.0
mA
mA output
Last measured value timeout
0.00 sec
Tertiary variable
Mass flow
Frequency factor
1,000.00 Hz
.00091 –
10,000.00 Hz
Rate factor
16,666.66992
g/s
Frequency pulse width
0 (50% duty
cycle)
0.01 – 655.35
millisec
Scaling method
Freq=Flow
Frequency fault action
Downscale
Frequency fault level – upscale
15,000 Hz
10.0 – 15,000
Hz
Frequency output polarity
Active high
Frequency
output
Last measured value timeout
0.0 seconds
0.0 – 60.0 sec
Configuration and Use Manual
205
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...