Quick start with the display
Lower Range Value
specifies the value of
Mass Flow Rate
to be represented by an output level of
0 mA or 4 mA.
a.
Activate
Scroll
until
AO 1 4 mA
or
AO 1 0 mA
appears on the display, then activate
Select
.
b.
Define each character in
Lower Range Value
, including the sign.
Use
Select
to highlight a specific character. Use
Scroll
to set the value of the character.
c.
When you have set all characters as desired, activate
Scroll
and
Select
simultaneously until
SAVE/YES?
flashes on the display, then activate
Select
to write the value to transmitter memory.
5.
Set
Upper Range Value
(URV).
Upper Range Value
specifies the value of
Mass Flow Rate
to be represented by an output level of 20 mA.
a.
Activate
Scroll
until
AO 1 20 mA
appears on the display, then activate
Select
.
b.
Define each character in
Upper Range Value
, including the sign.
Use
Select
to highlight a specific character. Use
Scroll
to set the value of the character.
c.
When you have set all characters as desired, activate
Scroll
and
Select
simultaneously until
SAVE/YES?
flashes on the display, then activate
Select
to write the value to transmitter memory.
6.
Activate
Scroll
until
AO EXIT
appears on the display, then activate
Select
.
7.
Return the display to normal operation (displaying process data).
a.
Activate
Scroll
until
IO EXIT
appears on the display, then activate
Select
.
b.
Activate
Scroll
until
CONFIG EXIT
appears on the display, then activate
Select
.
c.
Activate
Scroll
until
OFF-LINE EXIT
appears on the display, then activate
Select
.
d.
Activate
Scroll
until
EXIT
appears on the display, then activate
Select
.
Tip
Another way to exit the off-line menu is to rely on the time-out feature. If you have not used
the off-line menu for approximately 60 seconds, the display will automatically return to normal
operation.
2.4
Perform a loop test
A loop test is a way to verify that the transmitter and the receiving device are communicating properly.
A loop test also helps you know whether you need to trim mA outputs. Performing a loop test is not
a required procedure. However, Micro Motion recommends performing a loop test for every input or
output available on your transmitter.
Procedure
1.
Test the mA output.
a.
Choose
OFFLINE MAINT→SIM→AO1 SIM
, and select
SET 4 MA
or another mA output value.
Dots traverse the display while the output is fixed.
b.
Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct
the discrepancy by trimming the output.
Configuration and Use Manual
21
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...