Using the transmitter display
a.
With the cursor flashing on the first digit, activate
Scroll
until the correct digit is displayed,
then activate
Select
.
b.
Repeat this process for the second, third, and fourth digits.
Tip
If you do not know the correct value for
Off-Line Password
, wait 30 seconds. The password screen
will time out automatically and you will be returned to the previous screen.
4.
If
Scroll
flashes on the display, activate the
Scroll
optical switch, then the
Select
optical switch, and
then the
Scroll
optical switch again.
The display will prompt you through this sequence. The
Scroll-Select-Scroll
sequence is designed to
protect the display from accidental activation of the off-line menu. It is not designed as a security
measure.
5.
To exit a display menu and return to a higher-level menu:
• Activate
Scroll
until the
EXIT
option is displayed, then activate
Select
.
• If the
EXIT
option is not available, activate
Scroll
and
Select
simultaneously and hold until
the screen returns to the previous display.
6.
To exit the display menu system, you can use either of the following methods:
• Exit each menu separately, working your way back to the top of the menu system.
• Wait until the display times out and returns to displaying process variable data.
C.2.1
Optical switches
The transmitter has two optical switches:
Scroll
and
Select
. To activate an optical switch, block the light by
holding your thumb or finger in front of the opening.
If you are in a hazardous environment, do not remove the transmitter housing cover.
Removing the transmitter housing cover in a hazardous environment can cause an
explosion or other damage. You can operate the optical switches through the lens.
The optical switch indicator lights up when the transmitter senses that an optical switch has been
activated. See Table C-1 .
Table C-1
Optical switch indicator and optical switch states
Optical switch indicator
State of optical switches
Solid red
Either the
Scroll
or the
Select
optical switch is activated.
Flashing red
Both optical switches are activated.
C.2.2
Enter a floating-point value
The display allows you to enter a maximum of 8 characters, including the sign. The decimal point is not
counted as a character. Exponential notation is used to enter values that require more than 8 characters.
216
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...