Configure device options and preferences
Status alarms, default severity settings, and related information are listed in Table 6-1.
Table 6-1
Status alarms and
Status Alarm Severity
Alarm
code
Status message
Default
severity
Notes
Configurable?
A003
Sensor failure
Fault
Yes
A004
Temperature sensor failure
Fault
No
A005
Input overrange
Fault
Yes
A006
Transmitter not configured
Fault
Yes
A008
Density overrange
Fault
Yes
A009
Transmitter initializing/warming
up
Fault
Yes
A010
Calibration failure
Fault
No
A011
Calibration too low
Fault
Yes
A012
Calibration too high
Fault
Yes
A013
Zero too noisy
Fault
Yes
A014
Transmitter failed
Fault
No
A016
Line temperature out-of-range
Fault
Yes
A017
Meter RTD temperature
out-of-range
Fault
Yes
A018
EEPROM checksum error
Fault
No
A019
RAM or ROM test error
Fault
No
A020
Calibration factors unentered
Fault
Yes
A021
Incorrect sensor type
Fault
No
A027
Security breach
Fault
No
A100
Primary mA output saturated
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A101
Primary mA output fixed
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
A102
Drive overrange
Informational
Yes
A104
Calibration in progress
Informational
Can be set to either
Informational
or
Ignore
, but cannot be set to
Fault
.
Yes
110
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...