Integrate the meter with the control system
Table 7-7
Options for
Frequency Output Fault Action
Code
ProLink II
Field Communicator
Frequency output behavior
Upscale
Upscale
Goes to configured
Upscale
value:
•
Range: 10 Hz to 15000 Hz
•
Default: 15000 Hz
Downscale
Downscale
0 Hz
Internal Zero
Intrnl Zero
0 Hz
None
(default)
None
(default)
Tracks data for the assigned process variable
If you set
mA Output Fault Action
or
Frequency Output Fault Action
to
None
, be sure to set
Digital
Communications Fault Action
to
None
. If you do not, the output will not report actual process data,
and this may result in measurement error or unintended consequences for your process.
If you set
Digital Communications Fault Action
to
NAN
, you cannot set
mA Output Fault Action
or
Frequency Output Fault Action
to
None
. If you try to do this, the transmitter will not accept the
configuration.
7.2.5
Configure
Frequency Output Power Source
Display
OFF-LINE MAINT→OFF-LINE CONFG→IO→FO→POWER
ProLink II
ProLink→Configuration→Frequency→Power Type
Field Communicator
Configure→Manual Setup→Inputs/Outputs→Frequency Output→FO Settings→Power Source
Use
Frequency Output Power Source
to set the output power source for the frequency output. The power
configuration must match the wiring for the frequency output.
Procedure
Set
Frequency Output Power Source
as desired.
Option
Description
Internal
The output is powered by the transmitter
External
The output is powered by an external power
source.
7.3
Configure the discrete output
Display
OFF-LINE MAINT→OFF-LINE CONFG→IO→DO
ProLink II
ProLink→Configuration→Discrete Output
Field Communicator
Configure→Manual Setup→Inputs/Outputs→Discrete Output
The discrete output is used to report specific flowmeter or process conditions. The discrete output
parameters control which condition is reported and how it is reported.
Configuration and Use Manual
129
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...