Configure device options and preferences
This function is accessed from the totalizer value on the display. If the appropriate totalizer is not
configured as a display variable, the operator will not be able to reset it.
6.2.3
Enable or disable the
Acknowledge All Alarms
display command
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→ALARM
ProLink II
ProLink→Configuration→Display→Display Options→Display Ack All Alarms
Field Communicator
Not available
You can control whether or not the operator will be able to use a single command to acknowledge
all alarms from the display.
Procedure
Enable or disable
Acknowledge All Alarms
as desired.
Option
Description
Enabled
(default)
Operators can use a single display command to
acknowledge all alarms at once.
Disabled
Operators cannot acknowledge all alarms at once.
They must acknowledge alarms individually.
Note
To acknowledge alarms from the display, the operator must have access to the alarm menu, whether
Acknowledge All Alarms
is enabled or disabled.
6.3
Configure security for the display menus
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→OFFLN
ProLink II
ProLink→Configuration→Display→Display Options→Display Offline Menu
Field Communicator
Not available
You can control operator access to different sections of the display off-line menu, and you can control
whether or not a password is required at specific entry points.
Procedure
1.
To control operator access to the maintenance section of the off-line menu, enable or disable
Off-Line Menu
.
Configuration and Use Manual
105
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...