Configure device options and preferences
Display variable
Process variable assignment
Display Variable 6
Temperature
Display Variable 7
External pressure
Display Variable 8
Mass flow
Display Variable 9
None
Display Variable 10
None
Display Variable 11
None
Display Variable 12
None
Display Variable 13
None
Display Variable 14
None
Display Variable 15
None
6.1.3
Configure the precision of process variables shown on the
display
Display
Not available
ProLink II
ProLink→Configuration→Display→Display Precision
Field Communicator
Configure→Manual Setup→Display→Set Up Decimal Places→For Process Variables
For each process variable,
Display Precision
controls the number of digits to the right of the decimal place
that are shown on the display. You can set
Display Precision
independently for each process variable.
Display Precision
does not affect the value of the process variable reported via other methods or used
in calculations.
Procedure
1.
Select a process variable.
2.
Set
Display Precision
to the precision to be applied when this process variable is shown on the display.
For temperature and density process variables, the default value is 2. For all other process
variables, the default value is 4. The range is 0 to 5.
Tip
The lower the precision, the larger a process change must be in order to be reflected in the
displayed value. Do not set
Display Precision
too low or too high to be useful.
6.1.4
Configure the refresh rate of data shown on the display
Display
OFF-LINE MAINT→OFF-LINE CONFG→DISPLAY→RATE
ProLink II
ProLink→Configuration→Display→Display Options→Update Period
Field Communicator
Configure→Manual Setup→Display→Update Period
Update Period
controls how often the display is refreshed with current data.
Configuration and Use Manual
101
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...