Configure device options and preferences
Display Language
controls the language used for process data and menus on the display. Different
languages are available, depending on your transmitter model and version.
Procedure
Set
Display Language
to the desired option.
6.1.2
Configure the process variables that are shown on the display
Display
Not available
ProLink II
ProLink→Configuration→Display→Display Var X
Field Communicator
Configure→Manual Setup→Display→Set Up Display Variables
Display Variables
controls the process variables shown on the display. The display can scroll through up to
15 process variables in any order. You can configure the process variables to be displayed and the
order in which they will appear. You can repeat variables, and you can leave slots empty.
Restrictions
• You cannot set
Display Variable 1
to
None
.
Display Variable 1
must always be set to a process variable.
• If you have fixed
Display Variable 1
to the primary mA output, you cannot change the setting of
Display Variable 1
using this method. To change the setting of
Display Variable 1
, you must change the
configuration of
mA Output Process Variable
for the primary mA output.
Note
If you have configured a volume process variable as a display variable, and you subsequently change
the setting of
Volume Flow Type
, the display variable is automatically changed to the equivalent process
variable. For example, if
Display Variable 2
was set to
Volume Flow Rate
, it will be changed to
Gas Standard
Volume Flow Rate
.
Procedure
1.
Select a slot.
2.
Set
Display Variable
to the variable that you want to appear in that slot.
t
Example: Display variable configuration
Display variable
Process variable assignment
Display Variable 1
Mass flow
Display Variable 2
Mass totalizer
Display Variable 3
Volume flow
Display Variable 4
Volume totalizer
Display Variable 5
Density
100
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...