Integrate the meter with the control system
4.
Set a value for the setpoint (
Setpoint A
).
5.
(Optional) Configure a discrete output to switch states according to event status.
7.7.2
Configure an enhanced event
Display
Not available
ProLink II
ProLink→Configuration→Discrete Events
Field Communicator
Configure→Alert Setup→Discrete Events→Discrete Events 1–5
An “enhanced” event is used to perform specific transmitter actions if the event occurs. An enhanced
event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI)
or below (LO) a user-defined setpoint, or in range (IN) or out of range (OUT) with respect to two
user-defined setpoints. You can define up to five enhanced events. For each enhanced event, you can
assign one or more actions that the transmitter will perform if the enhanced event occurs.
Procedure
1.
Select
Event 1
,
Event 2
,
Event 3
,
Event 4
, or
Event 5
from
Event Name
.
2.
Specify
Event Type
.
HI
The event will occur if the value of the assigned process variable (
x
) is
greater than the setpoint (
Setpoint A
), endpoint not included.
x
> A
LO
The event will occur if the value of the assigned process variable (
x
) is less
than the setpoint (
Setpoint A
), endpoint not included.
x
< A
IN
The event will occur if the value of the assigned process variable (
x
) is “in
range,” i.e., between
Setpoint A
and
Setpoint B
, endpoints included.
A
≤
x
≤
B
OUT
The event will occur if the value of the assigned process variable (
x
) is “out of
range,” i.e., less than
Setpoint A
or greater than
Setpoint B
, endpoints included.
x
≤
A or
x
≥
B
3.
Assign a process variable to the event.
4.
Set values for the required setpoints.
• For HI or LO events, set
Setpoint A
.
• For IN or OUT events, set
Setpoint A
and
Setpoint B
.
Configuration and Use Manual
147
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...