Troubleshooting
Table 10-2
Status alarms and recommended actions
continued
Alarm code
Description
Cause
Recommended actions
A105
Slug flow
The density has
exceeded the
user-defined slug
(density) limits.
•
Check for slug flow. See Section 10.24.
A106
Burst mode enabled
The device is in HART
burst mode.
•
No action required.
•
If desired, you can reconfigure the alarm
severity level to
Ignore
.
A107
Power reset occurred
The transmitter has
been restarted.
•
No action required.
•
If desired, you can reconfigure the alarm
severity level to
Ignore
.
A108
Event 1 triggered
•
No action required.
•
Review event configuration if you believe
the event was triggered erroneously.
A109
Event 2 triggered
•
No action required.
•
Review event configuration if you believe
the event was triggered erroneously.
A110
Frequency output
saturated
The calculated
frequency output is
outside of the linear
range.
•
Check the frequency output scaling. See
Section 10.20 .
•
Check process conditions. Actual
conditions may be outside of the normally
expected conditions for which the output
is configured.
•
Verify process conditions, checking
especially for air in the flow tubes, tubes
not filled, foreign material in the tubes, or
coating in the tubes.
•
Verify that the measurement units are
configured correctly for your application.
•
Purge the flow tubes.
A111
Frequency output fixed
User has fixed the
frequency output.
•
Check that the transmitter is in loop test
mode.
•
Check that the output has been fixed via
digital communication.
182
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...