Integrate the meter with the control system
Important
To minimize configuration requirements, the 9739 MVD transmitter uses an auto-detection scheme
when responding to a connection request. With this auto-detect feature, you do not need to enter some
HART communication parameters.
Procedure
Set
HART Address
to a value between
0
and
15
.
HART Address
must be unique on the network. The default address (0) is typically used unless you are
in a multidrop environment.
Tip
Devices using HART protocol to communicate with the transmitter may use either
HART Address
or
HART
Tag
(
Software Tag
) to identify the transmitter. You may configure either or both, as required by your other
HART devices.
7.6.3
Configure Modbus/RS-485 communications
Display
Not available
ProLink II
ProLink→Configuration→Device→Digital Comm Setting
Field Communicator
Configure→Manual Setup→Inputs/Outputs→Communications→Setup RS-485 Port
Modbus/RS-485 communications parameters control Modbus communication with the transmitter's
RS-485 terminals.
Modbus/RS-485 communications parameters include:
•
Modbus Address
(
Slave Address
)
•
Disable Modbus ASCII
•
Floating-Point Byte Order
•
Additional Communications Response Delay
Important
To minimize configuration requirements, the 9739 MVD transmitter uses an auto-detection scheme
when responding to a connection request. With this auto-detect feature, you do not need to enter some
Modbus communication parameters.
Restriction
To configure
Floating-Point Byte Order
or
Additional Communications Response Delay
, you must use ProLink II.
Procedure
1.
Set
Disable Modbus ASCII
as desired.
Configuration and Use Manual
143
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...