Configure process measurement
Procedure
1.
Specify
Sensor Type
.
•
Straight-tube
(T-Series)
•
Curved-tube
(all sensors except T-Series)
2.
Set the flow characterization parameters. Be sure to include all decimal points.
• For straight-tube sensors, set
FCF
(
Flow Cal
or
Flow Calibration Factor
),
FTG
, and
FFQ
.
• For curved-tube sensors, set
Flow Cal
(
Flow Calibration Factor
).
3.
Set the density characterization parameters.
• For straight-tube sensors, set
D1
,
D2
,
DT
,
DTG
,
K1
,
K2
,
FD
,
DFQ1
, and
DFQ2
.
• For curved-tube sensors, set
D1
,
D2
,
TC
,
K1
,
K2
, and
FD
. (
TC
is sometimes shown as
DT
.)
5.1.1
Sources and formats for characterization parameters
Different sensor tags display characterization parameters differently, and older sensors may not have all
the required parameters on the tag.
Sample sensor tags
Sample sensor tags are shown in the following illustrations:
• Figure 5-1 : Older curved-tube sensors (all sensors except T-Series)
• Figure 5-2 : Newer curved-tube sensors (all sensors except T-Series)
• Figure 5-3 : Older straight-tube sensors (T-Series)
• Figure 5-4 : Newer straight-tube sensors (T-Series)
Figure 5-1
Tag on older curved-tube sensors (all sensors except T-Series)
62
Micro Motion 9739 MVD Transmitters
Содержание Micro Motion 9739 MVD
Страница 6: ......
Страница 8: ......
Страница 30: ...Quick start with the display OFFLINE MAINT RESTORE ZERO RESTORE YES 24 Micro Motion 9739 MVD Transmitters ...
Страница 54: ......
Страница 66: ......
Страница 104: ......
Страница 122: ......
Страница 158: ......
Страница 160: ......
Страница 170: ......
Страница 182: ......
Страница 206: ......
Страница 207: ...Appendices and reference ...
Страница 208: ......
Страница 218: ......
Страница 230: ...Using the transmitter display Figure C 4 Offline menu version information 224 Micro Motion 9739 MVD Transmitters ...
Страница 231: ...Using the transmitter display Figure C 5 Offline menu configuration units and I O Configuration and Use Manual 225 ...
Страница 233: ...Using the transmitter display Figure C 7 Offline menu configuration display Configuration and Use Manual 227 ...
Страница 234: ...Using the transmitter display Figure C 8 Offline menu Simulation loop testing 228 Micro Motion 9739 MVD Transmitters ...
Страница 236: ...Using the transmitter display Figure C 10 Offline menu Zero 230 Micro Motion 9739 MVD Transmitters ...
Страница 239: ...Using ProLink II with the 9739 MVD transmitter Figure D 2 Configuration menu Configuration and Use Manual 233 ...
Страница 241: ...Using ProLink II with the 9739 MVD transmitter Figure D 4 Configuration menu continued Configuration and Use Manual 235 ...
Страница 242: ......
Страница 245: ...Using the Field Communicator with the 9739 MVD transmitter Figure E 2 Overview menu Configuration and Use Manual 239 ...