SIMS GO
59
Scan Overlap
The scan overlap field shows how much overlap between each track is applied on the plate. The
overlap values range from 0mm to 100mm. Altering this value realigns the tracks on the plate to
reflect the new overlap value. This also affect the number of tracks used to cover the plate.
Figure 69: Overlap value of 0mm
Figure 70: Overlap value of 50mm
Shell Offset
The shell offset field is available when scanning Annular plates and refers to the inward distance of
the curved scan from the tank shell wall. The angle of the curve is maintained.
Measuring Shell Offset
Position the scanner on the annular plate for the desired curved scan. Measure the perpendicular
distance from the notch indicating MFL sensor location on the side black casing surface of the
scanning head to the shell wall. If the scanner is against the shell wall the shell offset is zero.
Figure 71: Shell offset measurement
Figure 72: Scanning head side black casing notch
Creating a Plate
When all the plate set up information has been entered and the calibration matches the plate and
coating thickness you can create the plate. Progressing to plate view automatically creates the plate
in the inspection.
When returning to plate setup a created plate can be identified by a green circle with created status
appears below the plate image. For plates that have not yet been created an orange circle with new
status appears below the plate image.
Figure 73: Created plate status with green circle
Figure 74: New plate status with orange circle
If the plate shows a new status, all the parameters can be edited. As soon as the plate is created the
parameters become greyed out, and only deleting the plate will allow modifying the plate with new
parameters.
Содержание Floormap X
Страница 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Страница 10: ...10 Chapter 1 Precautions and conventions...
Страница 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Страница 19: ...19 Chapter 2 FloormapX System Overview...
Страница 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Страница 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Страница 69: ...69 Chapter 5 SIMS PRO...
Страница 98: ...98 Chapter 6 Troubleshooting and Maintenance...