SIMS GO
61
Add Indication
The add indication button allows a manual indication to be added onto the plate. All added
indications are automatically included into the indication list.
Display Threshold
The display threshold button allows the minimum estimated percentage level value from which
indications should be displayed from in plate view and scan view to be set.
Add Custom Scan
The add custom scan button allows a new scan to be added to any part of the plate and made in
any direction or angle on the plate. Once add custom scan is selected an option tab for the scan
setup becomes available on the right-hand side of the screen.
Figure 77: Plate view update for add custom scan
The orientation for an added scan on a plate can be one set to one of the following:
•
Angled.
•
Vertical.
•
Horizontal.
A start (x,y) and end (x,y) reference co-ordinates for the added scan are determined by measuring
from the reference corner of the plate to the chosen reference point of the scanner, either the left-
hand or right-hand side of the scanning bridge.
To add a custom scan:
1.
Position the scanner on the plate in preparation for the scan, then click
Add Custom Scan
.
2.
Select the required scan orientation.
3.
Place a finger or stylus on the screen and draw the scan on the plate.
NOTE
Use the circular anchors to alter the scans position on the plate.
4.
Select the required scanner reference side, either Left of Right of the scanning bridge (considered
when standing behind the FloormapX), to measurements to from the plate reference.
5.
Edit the start (x,y) and end (x,y) reference coordinates of the scan to match the measurements
from the plate reference.
NOTE
For vertical and horizontal scans, the end coordinates are not a requirement as the scan
direction is set and the length of the scan is determined by the distance travelled by the
Содержание Floormap X
Страница 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Страница 10: ...10 Chapter 1 Precautions and conventions...
Страница 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Страница 19: ...19 Chapter 2 FloormapX System Overview...
Страница 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Страница 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Страница 69: ...69 Chapter 5 SIMS PRO...
Страница 98: ...98 Chapter 6 Troubleshooting and Maintenance...