FloormapX System Overview
24
Cable Ports
The color-coded ports at either side of the lower mainframe are for connecting the mainframe to the
scanning head. When viewing the system from the front the left-hand side port is colored orange and
is for the scanning head motorization. The right-hand side port is colored blue and is for the probe
cables.
NOTE
To avoid damage when assembling the FloormapX please ensure cables are attached after
the modules have been correctly assembled and verified as secure.
Reposition wheels
Located at the bottom of the mainframe are the reposition wheels. These are used for disengaging
the FloormapX scanning head from the inspection for maneuvering and positioning.
Emergency stop
When activated the emergency stop prohibits the drive motors, the magnet motors and steering
motors functionality. It does not shutdown the power to the entire system. To activate press and the
switch latches. To reset rotate clockwise.
NOTE
If the system is not operating as expected then please ensure the Emergency Stop has not
been accidentally pressed, especially when assembling the FloormapX from the component
modules.
Laser and rear light module
Located at the rear of the mainframe is the rear light and laser module. When the system is powered
on the rear lighting is always on and at a constant brightness. The red laser line can be switched on
and off using the laser guide button on the user control. The specifications for the laser fitted to the
scanner are:
•
Laser Class: 2.
•
Wavelength: 650nm.
•
Maximum output power: 1mW.
•
Emission type: Continuous wave.
•
Maintenance: If required.
Figure 9: Laser and rear light module
NOTE
The user control light button does not control the rear lighting.
WARNING LASER
Do not deliberately look or stare into the laser beam. Avoid accidental exposure to eyes.
Lateral tilt knob
Located at the rear of the mainframe is the lateral tilt know. This allows the mainframe to lean either
left or right so the scanning head can get up close to the tanks shell and other restricted areas.
To lean the mainframe, unscrew the lateral tilt knob and pull to disengage the plunger. Lean the
mainframe in the desired direction until the plunger engages then screw the lateral tilt knob.
Содержание Floormap X
Страница 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Страница 10: ...10 Chapter 1 Precautions and conventions...
Страница 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Страница 19: ...19 Chapter 2 FloormapX System Overview...
Страница 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Страница 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Страница 69: ...69 Chapter 5 SIMS PRO...
Страница 98: ...98 Chapter 6 Troubleshooting and Maintenance...